Model C300 CO2 Module - SPECIFICATIONS | |
Carbon Dioxide Monitoring: | |
Mode of Sampling Principle of Operation |
Sidestream Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts |
Initialization Time | Capnogram displayed in less than 20 seconds, At an ambient temperature of 25°C, full specifications within 2 minutes |
CO2 Measurement Range |
0 to 150 mmHg, 0 to 19.7% 0 to 20 kPa (Barometric Pressure supplied by module and mmHg reading is auto-calculated ) |
CO2 Calculation Method | BTPS (Body Temperature Pressure Saturated) |
CO2 Response Time | <4 seconds -includes transport time and rise time |
CO2 Resolution | 0.1 mmHg 0 to 69 mmHg 0.25 mmHg 70 to 150 mmHg |
CO2 Accuracy* | 0 -40 mmHg ± 2 mmHg 41 -150 mmHg ±8% of reading Above 80 breath per minute ± 10% of reading * NOTE: Gas temperature at 25°C. |
CO2 Stability | Short /long Term Drift: Drift shall not exceed 1 mmHg maximum. Module will auto run zero calibration when drift exceed 1 mmHg. |
CO2 Noise | RMS noise of the module shall be less than or equal to 0.25 mmHg at 5% CO2 |
Sampling Rate | 100 Hz |
RespirationRate Range | 2 to 150 breaths per minute (BPM) |
Respiration Rate Accuracy |
± 1 breath |
Calibration ETCO2 Calculation Method: |
No routine user calibration required. Peak of the expired CO2 waveform |
Inspired CO2 Measurement |
Range: 3 to 50 mmHg Method: lowest reading of the CO2 waveform in the previous breath |
Compensations | need compensations for: Expired O2, Balance gas (N2, N2O, He) and Anesthetic Agents Compensations for barometric pressure has been finished by module itself. |